Cardio-TRAP® Base Board Installation
Step 1
·Lay back the mattress of the cath table and thoroughly clean the surface of the cath table with alcohol to provide a clean surface for Velcro adhesion.
·Wipe down the Velcro side of the Base Board with a sterile wipe. Do not remove the Velcro from the Base Board.
Step 2
·Locate the head edge of Base Board 14 inches from head edge of cath table. (You might use a piece of tape to mark the 14 inch mark on the table.)
Step 3
·Insert a house, hospital or car key in each of the two smallest slits on the right side of the Base Board.
·Butt the lower protruding end of each key against the table's edge.
·Insure that the head edge of the Base Board is still 14" from the head edge of the cath table.
Step 4
·Standing on the left side of the table, lift the left side of the Base Board up about 2 inches and peel off the plastic backing from the Velcro. Before sticking the Velcro to the table make sure both keys are touching the edge of the table on the right side (it may help if an assistant is gently pushing against the Base Board's edge on the right side of the table) and make sure the head edge of the Base Board is 14 inches from the head edge of the table.
Lower the left side of the Base Board so that it sticks to the table top.
·Press Base Board firmly over the Velcro to properly adhere it to the table.
Step 5
·Remove both keys used in Step 3 and lift the right side of the Base Board about 2 inches.
·Peel off the plastic backing from the Velcro and stick the right side of the Base Board to the table.
·Press Base Board firmly over the Velcro to properly adhere it to the table.
Step 6
·You may add a Velcro strip (supplied) to the center of the upper surface of the Base Board, between the mattress and the Base Board, to keep the mattress held in position on the Base Board .